280
Chapter 5 Error Messages
Execution Errors
5
-221 Settings conflict;
not able to modulate arb waveforms, modulation turned off
The function generator cannot generate a modulated waveform using the
arbitrary waveform function. The selected modulation mode has been
turned off.
-221 Settings conflict;
not able to modulate ramp waveforms, modulation turned off
The function generator cannot generate a modulated waveform using the
ramp function. The selected modulation mode has been turned off.
-221 Settings conflict;
not able to sweep pulse, sweep turned off
The function generator cannot generate a sweep using the pulse
function. The sweep mode has been turned off.
-221 Settings conflict;
not able to modulate dc, modulation turned off
The function generator cannot generate a modulated waveform using the
dc voltage function. The selected modulation mode has been turned off.
-221 Settings conflict;
not able to sweep dc, sweep turned off
The function generator cannot generate a sweep using the dc voltage
function. The sweep mode has been turned off.
-221 Settings conflict;
not able to burst dc, burst turned off
The function generator cannot generate a burst using the dc voltage
function. The burst mode has been turned off.
-221 Settings conflict;
not able to burst pulse, burst turned off
The function generator cannot generate a burst using the pulse function.
The burst mode has been turned off.
-221 Settings conflict;
not able to burst arb waveforms, burst turned off
The function generator cannot generate a burst using the arbitrary
waveform function. The burst mode has been turned off.
33210A users guide.book Page 280 Wednesday, July 16, 2008 11:16 AM