Chapter 5 Error Messages
Execution Errors
5
-222 Data out of range;
pulse edge time limited by period; value clipped to upper limit
The specified edge time must fit within the existing period and width.
The function generator will limit the edge time as needed to
accommodate the existing period.
-222 Data out of range;
pulse width limited by period; value clipped to ...
The specified pulse width must be less than the difference between the
period and the edge time as shown below. The function generator will
adjust the pulse width as needed to accommodate the specified period.
Pulse Width <
Period (1.6 X Edge Time)
-222 Data out of range;
pulse duty cycle limited by period; value clipped to ...
The specified pulse duty cycle must conform to the following period and
edge time restriction. The function generator will adjust the pulse duty
cycle as needed to accommodate the specified period.
Duty Cycle <
100 X (Period (1.6 X Edge Time)) ÷ Period
-222 Data out of range;
large period limits minimum pulse width
For a pulse waveform, the minimum pulse width is 40 ns for periods up
to 10 seconds. For periods above 10 seconds, the minimum pulse width is
greater. See “Pulse Configuration Commands” in chapter 4 for further
information.
-222 Data out of range;
pulse edge time limited by width; value clipped to ...
The specified edge time must fit within the specified pulse width as
shown below. The function generator will adjust the edge time as needed
to accommodate the specified pulse width.
Edge Time <
0.625 X Pulse Width