289
Chapter 5  Error Messages
Execution Errors
4
5
-222 Data out of range; 
frequency in FM; value clipped to ...
This generic message indicates that the carrier frequency has been 
limited to a lower boundary as determined by the FM:DEV command. 
The carrier frequency must always be greater than or equal to the 
frequency deviation. 
-222 Data out of range; 
marker confined to sweep span; value clipped to ...
This generic message indicates that the specified marker frequency is 
outside the range of the start frequency and stop frequency. The marker 
frequency must be between the specified start frequency and stop 
frequency. If you attempt to set the marker frequency to a frequency 
not in this range, the function generator will automatically set the 
marker frequency equal to the start frequency or stop frequency 
(whichever is closer). This error is generated only when both the sweep 
mode and the marker frequency are enabled. 
-222 Data out of range; 
pulse width; value clipped to ...
This generic message indicates that the desired pulse width is limited to 
an upper or lower boundary which is dictated by the instrument 
hardware.
-222 Data out of range; 
pulse edge time; value clipped to ...
This generic message indicates that the desired edge time is limited to 
an upper or lower boundary which is dictated by the instrument 
hardware.
-222 Data out of range; 
FM deviation; value clipped to ...
This generic message indicates that the desired FM deviation is limited 
to a lower or upper boundary as set by the current function’s frequency.
-222 Data out of range; 
FM deviation limited by minimum frequency
The frequency deviation is limited to the lower limit (1 mHz).
33210A users guide.book  Page 289  Wednesday, July 16, 2008  11:16 AM