60  
Chapter 4 Calibration Procedures
Performance Verification Tests
4
Special Note: 
Amplitude and Flatness Verification Procedures
Measurements made during the AC Amplitude (high-impedance) 
Verification procedure (see page 63) are used as reference measurements 
in the flatness verification procedures (beginning on page 64). Additional 
reference measurements and calculated references are used in the 
flatness verification procedures. Photo-copy and use the table on page 61 
to record these reference measurements and perform the calculations. 
The flatness verification procedures use both a DMM and a Power Meter 
to make the measurements. To correct the difference between the DMM 
and Power Meter measurements, the Power Meter reference 
measurement level is adjusted to set the 0.00 dB level to the DMM 
measurement made at 1 kHz. The flatness error of the DMM at 100 KHz 
is used to set the required 0.00 dB reference. 
The instrument internally corrects the difference between the high-Z 
input of the DMM and the 50 Ω input of the Power Meter when setting 
the output level. 
The reference measurements must also be converted from Vrms (made 
by the DMM) to dBm (made by the Power Meter). 
The equation used for the conversion from Vrms (High-Z) to dBm 
(at 50 Ω) is as follows:
Power (dBm) = 10 log(5.0 * V
rms
2
)
Flatness measurements for the –10 db, –20d B, and –30 dB attenuator 
ranges are verified as a part of the 0 dB verification procedure. No 
separate verification procedure is given for these ranges.