70
Chapter 4 Calibration Procedures
+20 dB Range Flatness Verification
4
4 Set the power meter offset to equal the 100kHz_20dB_offset value
previously calculated. This sets the power meter to directly read the
flatness error specification relative to 1 kHz. 100kHz_20dB_offset is
calculated on the Amplitude and Flatness Verification Worksheet.
5 Set the instrument to each output described in the table below and
measure the output amplitude with the power meter.
6 Compare the measured output to the test limits shown in the table.
Keysight 33210A
Measurement
Output Setup Function Amplitude Frequency Nominal Error
Q 50 Sine Wave +23.90 dBm 100.000 kHz 0 dB ±0.1 dB
50 Sine Wave +23.90 dBm 200.000 kHz 0 dB ±0.2 dB
50 Sine Wave +23.90 dBm 500.000 kHz 0 dB ±0.2 dB
50 Sine Wave +23.90 dBm 2.000 MHz 0 dB ±0.2 dB
50 Sine Wave +23.90 dBm 3.000 MHz 0 dB ±0.2 dB
Q 50 Sine Wave +23.90 dBm 4.000 MHz 0 dB ±0.2 dB
50 Sine Wave +23.90 dBm 5.000 MHz 0 dB ±0.2 dB
50 Sine Wave +23.90 dBm 8.000 MHz 0 dB ±0.3 dB
Q 50 Sine Wave +23.90 dBm 10.000 MHz 0 dB ±0.3 dB