Chapter 4 Calibration Procedures
AC Amplitude (high-impedance) Adjustment
4
4
3 Using the numeric keypad or knob, adjust the displayed voltage at each
setup to match the measured voltage. Select ENTER VALUE. (Entered
values are rounded to the nearest 100 µV).
4 After performing setup 28:
a. If your calibration procedures require you to verify the adjustment
just made, exit the calibration menu and perform “AC Amplitude
(high-impedance) Verification”, on page 63.
b. If you are making all the adjustments and then verifying th
e
i
nstrument’s performance, continue with the next procedure in
this chapter.