90
Chapter 4 Calibration Procedures
+20 dB Range Flatness Adjustment
4
* Constants are stored after completing this setup.
Nominal Signal
Setup Frequency Amplitude
63* 100 kHz 2.8 Vrms 22 dBm Flatness for 20 dB, Elliptical Filter
64* 200 kHz 2.8 Vrms 22 dBm Flatness for 20 dB, Elliptical Filter
65* 500 kHz 2.8 Vrms 22 dBm Flatness for 20 dB, Elliptical Filter
66* 1.5 MHz 2.8 Vrms 22 dBm Flatness for 20 dB, Elliptical Filter
67* 3.0 MHz 2.8 Vrms 22 dBm Flatness for 20 dB, Elliptical Filter
68* 4.0 MHz 2.8 Vrms 22 dBm Flatness for 20 dB, Elliptical Filter
69* 6.0 MHz 2.8 Vrms 22 dBm Flatness for 20 dB, Elliptical Filter
70* 8.0 MHz 2.8 Vrms 22 dBm Flatness for 20 dB, Elliptical Filter
71* 9.9 MHz 2.8 Vrms 22 dBm Flatness for 20 dB, Elliptical Filter
72* 200 kHz 2.8 Vrms 22 dBm Flatness for 20 dB, Linear Phase Filter
73* 500 kHz 2.8 Vrms 22 dBm Flatness for 20 dB, Linear Phase Filter
74* 1.5 MHz 2.8 Vrms 22 dBm Flatness for 20 dB, Linear Phase Filter
75* 3.0 MHz 2.8 Vrms 22 dBm Flatness for 20 dB, Linear Phase Filter