l CALCulate:SCALe:UNIT:STATe controls whether the unit string is displayed when the scaling function
is enabled.
l The instrument resets the units to the null string after a Factory Reset (*RST) or an Instrument Preset
(SYSTem:PRESet).
CALCulate:SCALe:UNIT:STATe {OFF|ON}
CALCulate:SCALe:UNIT:STATe?
Enables or disables displaying the unit string with measurements on the front panel when the scaling func-
tion is enabled.
Parameter Typical Return
{ON|1|OFF|0}. Default: OFF. 0(OFF)or1(ON)
Enable scaling with a gain of 100 and an offset of 5, and display "PSI" with measurements on the front
panel.
CALC:SCAL:FUNC SCAL
CALC:SCAL:GAIN 100
CALC:SCAL:OFFS 5
CALC:SCAL:UNIT "PSI"
CALC:SCAL:UNIT:STAT ON
CALC:SCAL:STAT ON
l Applies only to the 34465A and 34470A.
l CALCulate:SCALe:UNIT specifies the unit string to be displayed.
l The instrument disables displaying the unit string after a Factory Reset (*RST) or an Instrument Preset
(SYSTem:PRESet).
SCPI Programming Reference
228
Keysight Truevolt Series Operating and Service Guide