CALCulate:SCALe:OFFSet {<offset>|MIN|MAX|DEF}
CALCulate:SCALe:OFFSet? [{MIN|MAX|DEF}]
Stores an offset value B for the scaling function in CALCulate:SCALe:FUNCtion.
Parameter Typical Return
-1.0E+15 to -1.0E-15, 0.0, +1.0E-15 to 1.0E+15. Default: 0.0. +5.00000000E+00
Enable Mx-B scaling with a gain of 100 and an offset of 5.
CALC:SCAL:FUNC SCAL
CALC:SCAL:GAIN 100
CALC:SCAL:OFFS 5
CALC:SCAL:STAT ON
l Applies only to the 34465A and 34470A.
l The instrument sets the offset value to 0.0 after a Factory Reset (*RST) or Instrument Preset
(SYSTem:PRESet).
CALCulate:SCALe:REFerence {<reference>|MIN|MAX|DEF}
CALCulate:SCALe:REFerence? [{MIN|MAX|DEF}]
Stores a reference value for the PCT function in CALCulate:SCALe:FUNCtion.
Parameter Typical Return
-1.0E+15 to -1.0E-15, 0.0, +1.0E-15 to 1.0E+15. Default: 0.0. +3.00000000E+02
Enable PCT scaling with a 100V reference.
CALC:SCAL:REF 100
CALC:SCAL:FUNC PCT
CALC:SCAL:STAT ON
l Applies only to the 34465A and 34470A.
l Specifying a reference value disables automatic reference selection (CALCu-
late:SCALe:REFerence:AUTO OFF).
l The instrument sets the reference value to 0.0 with automatic reference selection enabled after a Fact-
ory Reset (*RST) or Instrument Preset (SYSTem:PRESet), or a change in measurement function.
Keysight Truevolt Series Operating and Service Guide