Capacitance Verification (Optional Verification Test)
This optional procedure verifies capacitance accuracy.
1.
Make sure you have read Test Considerations.
2.
Select the Capacitance measurement function.
3.
Connect the capacitance calibrator to the front input terminals and select the front input terminals
with the Front/Rear switch.
4.
Select each range in the order shown below. Provide the indicated input.
5.
Compare measurement results to the test limits shown in the table. (Be certain to allow time for
source settling.)
Input
Capacitance Range
Error from
Nominal
1 μF 1 μF ± 0.005 μF
1 μF 10 μF ± 0.014 μF