Temperature (34460A and 34461A)
This topic applies only to 34460A/61 DMMs. For temperature measurements using the 34465A/70A, see
Temperature (34465A and 34470A).
This section describes how to configure 2-wire and 4-wire temperature measurements from the front
panel.
Step 1: Configure the test leads as shown.
2-wire Temperature:
4-wire Temperature:
Step 2: Press [Temp] on the front panel. The following menu appears.
Step 3: Press Probe and choose the probe type. If you choose to use an RTD, the menu will have a softkey
to specify the RTD's resistance at 0 degrees Celsius (R
0
).
Keysight Truevolt Series Operating and Service Guide