Calibration Adjustment Process
The following general procedure is recommended to complete a full calibration adjustment.
1.
Read Test Considerations.
2.
Perform the performance verification tests to characterize the instrument (incoming data).
3.
Unsecure the instrument for calibration (see Calibration Security).
4.
Perform the calibration procedures (see Calibration Procedures).
5.
Secure the instrument against calibration (see Calibration Security).
6.
Note the security code and calibration count in the instrument’s maintenance records.
7.
Perform the performance verification tests to verify the calibration.
Keysight Truevolt Series Operating and Service Guide
527
Calibration Adjustment Procedures