[SENSe:]CURRent[:DC]:SECondary {"OFF"|"CALCulate:DATA"|"CURRent:AC"|"PTPeak"}
[SENSe:]CURRent[:DC]:SECondary?
Selects the secondary measurement function for DC current measurements.
Parameter Typical Return
{"OFF"|"CALCulate:DATA"|"CURRent:AC"|"PTPeak"},
default "OFF"
"CALC:DATA"
Configure DC current measurements using the 1 A range with 1 mA resolution. Select peak-to-peak as the
secondary measurement. Make and retrieve measurements.
CONF:CURR:DC 1,0.001
CURR:DC:SEC "PTP"
READ?;DATA2?
Typical Response: +5.54141768E-06;+9.91000000E+3,+9.91000000E+3
l "CALCulate:DATA" - (Applies only to the 34465A and 34470A.) The measurement value before any
math operations are done (including NULL).
l "CURRent:AC" - The AC current measurement of the input signal. Applies only when using the instru-
ment from the front panel. Setting is ignored for measurements made from a remote interface. If you
require both AC and DC current measurements from remote, program each function separately, see
CONFigure:CURRent:{AC|DC}.
l "PTPeak" - The peak-to-peak, maximum, and minimum values of the input signal.
l This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
(SYSTem:PRESet).
Keysight Truevolt Series Operating and Service Guide
331
SCPI Programming Reference