[SENSe:]{FRESistance|RESistance}:SECondary {"OFF"|"CALCulate:DATA"}
[SENSe:]{FRESistance|RESistance}:SECondary?
Selects the secondary measurement function for all resistance measurements.
Parameter Typical Return
{"OFF"|"CALCulate:DATA"}, default "OFF" "CALC:DATA"
Configure resistance and secondary measurements. Make resistance measurement and retrieve secondary
measurement.
CONF:RES AUTO,MAX
RES:SEC "CALC:DATA"
READ?;DATA2?
Typical Response: +3.01566373E-10;,+9.91000000E+3
l "CALCulate:DATA" - (Applies only to the 34465A and 34470A.) The measurement value before any
math operations are done (including NULL).
l Use READ? or INITiate to start the measurement. Use [SENSe:]DATA2? to retrieve the secondary meas-
urement.
l This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
(SYSTem:PRESet).
SCPI Programming Reference
356
Keysight Truevolt Series Operating and Service Guide