[SENSe:]TEMPerature:SECondary {"OFF"|"CALCulate:DATA"|"SENSe:DATA"}
[SENSe:]TEMPerature:SECondary?
Selects the secondary measurement function for temperature measurements.
Parameter Typical Return
{"OFF"|"CALCulate:DATA"|"SENSe:DATA"}, default
"OFF"
"CALC:DATA"
Configure a 4-wire RTD measurement. Select the raw sensor value (ohms in this example) as the secondary
measurement. Make and retrieve measurements.
CONF:TEMP FRTD,85
TEMP:SEC "SENS:DATA"
READ?;DATA2?
Typical Response: -6.95327319E-01;+9.97282173E+01
l "CALCulate:DATA" - (Applies only to the 34465A and 34470A.) The measurement value before any
math operations are done (including NULL).
l SENSe:DATA is the raw sensor value; resistance for thermistor/RTD, voltage and reference tem-
perature for thermocouple (thermocouple measurements apply to the 34465A and 34470A only).
l Use READ? or INITiate to start the measurement.
l This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
(SYSTem:PRESet).
Keysight Truevolt Series Operating and Service Guide
363
SCPI Programming Reference