EasyManua.ls Logo

Keysight 34470A - Page 420

Keysight 34470A
583 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SYSTem:LOCK:REQuest?
Requests a lock of the current I/O interface. This allows you to lock the instrument's configuration or
cooperatively share the instrument with other computers.
Parameter Typical Return
(none) +0 (denied) or +1 (granted)
See
Interface Locking Example
l Lock requests can be nested; each request increases lock count by 1. For every request, you need a
release (SYSTem:LOCK:RELease) from the same I/O interface.
l Locks are handled at the I/O interface level (USB, LAN, etc.) and you are responsible for all coordin-
ation between threads and/or programs on that interface.
l When a request is granted, only I/O sessions from the present interface can change the instrument
state. You can only query the instrument state from other I/O interfaces.
l LAN sessions locks are automatically released when a LAN disconnect is detected.
l Granting a lock sets Bit 10 in the Standard Operation Register (STATus:OPERation:CONDition?). In
addition, the entire instrument front panel, including the Local key, is locked out while a lock is in
place ("Front panel locked." is displayed).
Keysight Truevolt Series Operating and Service Guide