1-4 Keysight B1500A Configuration and Connection Guide, Edition 6
B1500A Product Configuration
Product line up of the B1500A series
N1302A-200 GSWU
Guard switch unit (GSWU). Needs the SCUU to use the GSWU. Effective for the
accurate impedance measurements by connecting the guard lines between CMU high
and low near the DUT.
B1542A pulsed IV package
B1542A expands the capabilities of the B1500A to enable the ultra short pulsed IV
measurements with parametric characterization down to 10 ns pulse width for many
new device structures.
Table 1-1 Mainframe Specification
Table 1-2 SMU Module Variation
Available Slots 10
Ground Unit (GNDU) sink capability 4.2 A
Instrument control GPIB
External trigger inputs / outputs
1 BNC Trigger in, 1 BNC Trigger out, 8
Programmable Trigger In/Out
Networking 100 BASE-TX / 10 BASE-T LAN Port
Module
Key specification
Max.
Current
Max.
Voltage
Current
means.
Resolu-
tion
Voltage
means.
Resolu-
tion
Req’d
slots
HPSMU
(High Power SMU)
±1 A ±200 V
10 fA 2 V2
MPSMU
(Medium Power SMU)
±100 mA ±100 V
10 fA 0.5 V1
HRSMU
(High Resolution SMU)
±100 mA ±100 V
1 fA 0.5 V1
ASU
(Atto Sense and Switch Unit)
±100 mA ±100 V
0.1 fA 0.5 VNA
MCSMU
(50 s pulse Medium Current SMU)
±1 A ±30 V
10 pA 0.2 V1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight B1500A Series and is the answer not in the manual?

Keysight B1500A Series Specifications

General IconGeneral
BrandKeysight
ModelB1500A Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals