13
Record Playback (Opt 236)
Allows you to record SA traces and play them back at a
later time. You can change most SA settings during a
recording. Setting changes are recorded along with the
traces.
To record a new session:
Press Trace 6 then Record Playback
Then New Session, Recorder Player, then Record
Optionally set the following Recording Configuration:
- Record Time – limits record time.
- Record Source – choose where in the FieldFox data
flow that data is recorded. RawMeas is easiest.
- Record Interval – sets the amount of time to wait
between individual trace recordings.
- Segment Count – sets the number of traces to
record.
- FreqMaskTrigger – sets the power level which will
trigger a recording.
To play a session:
Press Recall Session select a session, press Open. Then
Recorder Player then Play.
Run/Hold can be used to Pause and Resume during
Record or Playback.
The maximum recording time is determined largely by
the amount of available memory. To maximize recording
time, select Storage Device then Internal or USB (flash
drive) or SD Card.