6-6 Site Preparation
6 Power Requirements
Basic Power Quality Survey
Power quality can affect system performance differently. The following procedure 
is intended as a guideline and may not be the total solution. Failure to meet these 
guidelines should serve as an indicator that a power quality consultant might be 
needed to conduct a more in-depth power quality survey.
1 With the system operating, measure harmonic distortion at the system-input 
connection. THD should be less than 5% and less than 3% for any single 
harmonic.
2 With the system operating, measure the ground-to-neutral voltage at the 
system-input connection; the voltage should be less than 4vp-p.
3 Turn the system power off and measure the line voltage at the system-input 
connection; record this reading. Turn the system on and begin operating mode. 
Measure the line voltage at the system-input connection again. The difference 
between the two measurements should be less than 2%.
Other problematic power qualities include momentary voltage interruptions, 
ground noise, and voltage spikes. A survey of these problems and others may 
require the services of a power quality expert with specialized equipment.