Replacing Assemblies 5
S-Series Oscilloscopes Service Guide 85
With the keyboard removed, you can now remove either the AutoProbe board or
the display/touch screen assembly.
1 To remove the AutoProbe board, remove the two M3 screws connecting the
board to the deck.
2 Slide the AutoProbe board off the deck to disengage the keyholes.