8 M8000 Series of BER Test Solutions Tips for Preventing Damage
1 Tips for Preventing Damage
Connecting & removing M8058A remote head
Tools required to connect/disconnect the cable
• 8710-2812 Torque wrench: 4-in-lb 6 mm
• 54964-03801 Counter wrench: 2 mm thick, dual-ended, fits 6 mm and
7 mm
Table 4 Steps to connect / disconnect the cable
Steps to be performed Reference image of the cable
Connecting remote head to Data Out on
module
Note that the pre-connected adapter
11921-60007 (1 mm male side) is part of the
remote head.
Use the 7 mm side of the 54964-03801
wrench to hold at position 1. Avoid turning
inner pin.
Use 8710-2812 wrench to torque at position
2.
Disconnecting remote head from Data Out on
module
Keep the adapter 11921-60007 (1 mm male
side) as part of remote head .
Use the 7 mm side of the 54964-03801
wrench to hold at position 1. Avoid turning
inner pin.
Use 8710-2812 wrench to loosen at position
2.