Test 1: DC Volts Verification Tests
Keysight M9181A Calibration and Verification Procedure 19
A “Q” in the Quick Check column ind icates a quick performance verification
test point. All measurements are taken at 32 NPLC.
Calibrator
Input Voltage
DMM Function DMM Range Quick Check
1 Year Error from
Nominal (± V)
0.2 DCV 0.2 Q 2.350E-05
0 DCV 0.2 Q 3.500E-06
–0.2 DCV 0.2 2.350E-05
2 DCV 2 1.900E-04
0 DCV 2 1.000E-05
–2 DCV 2 Q 1.900E-04
20 DCV 20 2.700E-03
0 DCV 20 3.000E-04
–20 DCV 20 2.700E-03
200 DCV 200 Q 2.100E-02
0 DCV 200 1.000E-03
–200 DCV 200 2.100E-02