Test 3: Ohms Gain Verification Test
22 Keysight M9181A Calibration and Verification Procedure
Calibrator
Input
Resistance (W)
DMM Function DMM Range (W) Quick Check
1 Year Error from
Nominal (± W)
04W W 200 6.000E-03
190 4W W 200 3.070E-02
04W W 2,000 4.000E-02
1,900 4W W 2,000 Q 2.680E-01
04W W 20,000 4.000E-01
19,000 4W W 20,000 2.680E+00
04W W 200,000 6.000E+00
190,000 4W W 200,000 4.400E+01
04W W 2,000,000 8.000E+01
1,900,000 4W W 2,000,000 Q 1.220E+03
04W W 20,000,000 6.000E+02
19,000,000 4W W 20,000,000 3.860E+04