EasyManua.ls Logo

Keysight M9181A - Page 35

Keysight M9181A
47 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration
Keysight M9181A Calibration and Verification Procedure 35
Calibration Steps
Calibrate Zeros (Calibration Step 1)
This step performs calibration on the “zero” measurement for the offset calibration
of DCV, 2W, and 4W.
This procedure uses a low-thermal shorting block across the input connectors.
1 Install the shorting block on the front panel voltage input terminals of the
DMM, as shown below.
2 Now Perform the calibration process (see “Calibration Process” on page 31).
Enter the calibration value of 0.00 for the calibration value (because the source
is 0 V).
SFP Cal Step Calibration Name Input Value VDC
1 Zeros 0.00 Low Thermal Short

Table of Contents

Other manuals for Keysight M9181A

Related product manuals