Calibration
Keysight M9181A Calibration and Verification Procedure 45
SFP Cal Step Function Range (W) Input Level Input Function 2W/4W W
43 (Skip for M9181A - Applies only to M9182A/83A)
44 Ohm-4W Gain 20 4W 19
45 Ohm-4W Gain 200 4W 190
46 Ohm-4W Gain 2,000 4W 1,900
47 Ohm-4W Gain 20,000 4W 19,000
48 Ohm-4W Gain 200,000 4W 190,000
49 Ohm-4W Gain 2,000,000 4W 1,900,000
50 Ohm-4W Gain 20,000,000 4W 19,000,000
51 (Skip for M9181A - Applies only to M9182A/83A)
52 (Skip for M9181A - Applies only to M9182A/83A)