Test 6: AC Volts Verification Test
30 Keysight M9182A/M9183A Calibration and Verification Procedure
Calib. source
input voltage
(V)
Calib. source
freq (Hz)
DMM
function
DMM
voltage range
(V)
Quick check
Error from nominal (± V)
24 hours 90 days 1 year
10.00000E-03 47.00000E+00 ACV 200.0000E-03 213.0E-06 234.0E-06 235.0E-06
10.00000E-03 50.00000E+03 ACV 200.0000E-03 285.0E-06 326.0E-06 367.0E-06
200.0000E-03 47.00000E+00 ACV 200.0000E-03 Q 360.0E-06 400.0E-06 420.0E-06
20.00000E-03 10.00000E+00 ACV 2.000000E+00 2.600E-03 2.820E-03 3.240E-03
2.000000E+00 47.00000E+00 ACV 2.000000E+00 2.000E-03 2.400E-03 2.600E-03
2.000000E+00 10.00000E+03 ACV 2.000000E+00 2.000E-03 2.400E-03 2.600E-03
20.00000E-03 50.00000E+03 ACV 2.000000E+00 1.264E-03 1.466E-03 1.670E-03
2.000000E+00 100.0000E+03 ACV 2.000000E+00 39.600E-03 41.800E-03 44.000E-03
20.0000E+00 10.00000E+03 ACV 20.00000E+00 22.00E-03 26.00E-03 28.00E-03
20.0000E+00 50.00000E+03 ACV 20.00000E+00 Q 54.00E-03 62.00E-03 70.00E-03
200.0000E+00 47.00000E+00 ACV 200.0000E+00 240.0E-03 300.0E-03 340.0E-03
160.0000E+00 50.00000E+03 ACV 200.0000E+00 588.0E-03 640.0E-03 712.0E-03
300.0000E+00 1.000000E+03 ACV 300.0000E+00 360.0E-03 450.0E-03 510.0E-03
3.0000E+00 50.00000E+03 ACV 300.0000E+00 218.4E-03 249.0E-03 309.6E-03
200.0000E-03 350.0000E+00
F ACV
[a]
20.00000E+00 13.86E-03 15.92E-03 18.00E-03
23.0000E+00 50.00000E+03
F ACV
[a]
20.00000E+00 89.30E-03 97.90E-03 106.50E-03
200.0000E-03 100.0000E+03
F ACV
[a]
20.00000E+00 34.00E-03 40.40E-03 44.80E-03
[a] Set the frequency filter to Fast (400 Hz).