You can test the TRAN_active bit in the Operation Status register to know when the transient trigger
system has returned to the idle state.
Front Panel Menu Reference SCPI Command
Not Available To query the TRAN_active bit (bit6):
STAT:OPER:COND?
If a bit value of 64 is returned in the query, the TRAN_active bit is true, and the transient action is NOT
complete. When the TRAN_active bit is false, the transient action is complete. Refer to Status Tutorial
for more information.
Generating Trigger-Out Signals
Each output channel can generate trigger signals that can be used by other output channels, or
routed to a pin on the digital port that has been configured as a trigger output (TOUT). Use the
following commands to program transient trigger signals that are generated when an output Step
occurs.
Front Panel Menu Reference SCPI Command
Use the Channel key to selectthe channel thatis the
trigger source. Select Transient\Step.
Check Enable Trig out. Then press Select.
To program channel 3’s step function to generate a
trigger signal, use
STEP:TOUT ON,(@3)
Keysight N6700C Operating and Service Guide 107
4 Operating the Power System