Current Commands
Current commands program the output current and current protection functions. The SOURce
keyword is optional in the following commands.
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <value>, (@<chanlist>)
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]? [MIN|MAX,] (@<chanlist>)
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude] <value>, (@<chanlist>)
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]? [MIN|MAX,] (@<chanlist>)
Sets the immediate current level and the triggered current level when the output is in constant current
mode. The triggered level is a stored value that will be programmed when a Step transient is
triggered. Units are in amperes. The maximum value depends on the current rating of the unit.
Parameter Typical Return
0 - maximum|MIN|MAX, *RST 0
The maximum value is 102% of the current rating of the module.
<current level>
Sets the output current level to 3 A: CURR 3, (@1)
Sets the triggered current level to 2 A: CURR:TRIG 2, (@1)
l The query returns the programmed current level in the form +n.nnnnnnE+nn for each channel spe-
cified. Multiple responses are separated by commas.
[SOURce:]CURRent:LIMit[:POSitive][:IMMediate][:AMPLitude] <value>, (@<chanlist>)
[SOURce:]CURRent:LIMit[:POSitive][:IMMediate][:AMPLitude]? [MIN|MAX,] (@<chanlist>)
[SOURce:]CURRent:LIMit:NEGative[:IMMediate][:AMPLitude] <value>, (@<chanlist>)
[SOURce:]CURRent:LIMit:NEGative[:IMMediate][:AMPLitude]? [MIN|MAX,] (@<chanlist>)
Sets the current limit when in voltage priority mode. Units are in amperes. The maximum value
depends on the current rating of the unit. The minimum value is the most negative value.
Parameter Typical Return
Positive: 0 to 102% of rating|MIN|MAX, *RST 102%
Negative: -102% of rating to 0|MIN|MAX, *RST -102%
<+ current limit>
<- current limit>
Sets the positive current limit to 2 A: CURR:LIM 2 (@1)
Sets the negative current limit to -2 A: CURR:LIM:NEG -2 (@1)
l MIN is the most negative value of the range. MAX is the least negative value of the range.
l The query returns the programmed current limit in the form +n.nnnnnnE+nn for each channel spe-
cified. Multiple responses are separated by commas.
l The negative current limit of Model N6783A-MFG is not programmable, but is fixed at -2A.