Step 3. Configure the final steps common to all Arbs. Refer to Configuring Steps Common to all Arbs at
the end of this section.
Configuring Exponential Arbs
Step 1. Select the Arb Type and Shape. Power and Resistance types apply to .
Front Panel Menu Reference SCPI Command
Select Transient\Arb\Function.
Select a Voltage, Current, Power, or Resistance Arb Type.
Then selectthe Exponential Shape. Then press Select.
To selecta current, voltage, power, or resistance exponential
Arb, use:
ARB:FUNC:TYPE VOLT|CURR|POW|RES,(@1)
ARB:FUNC:SHAP EXP,(@1)
Step 2. Configure the exponential parameters. The parameters apply to the selected Arb type
(Voltage, Current, Power, or Resistance).
Enter the values for the level before and after the exponential.
Enter the time before the exponential, the exponential time, and the time constant of the exponential.
Front Panel Menu Reference SCPI Command
Select Transient\Arb\Config\Exp\Level.
Enter the Start and End level in volts. Then press Select.
Select Transient\Arb\Config\Exp\Time.
Enter the Start, Time, and Time Constant (TC) in seconds.
Then press Select.
To configure a voltage step Arb, use:
ARB:VOLT:EXP:STAR 0, (@1)
ARB:VOLT:EXP:END 10, (@1)
ARB:VOLT:EXP:STAR:TIM 0.25,(@1)
ARB:VOLT:EXP:TIM 0.5,(@1)
ARB:VOLT:EXP:TCON 0.25,(@1)
Step 3. Configure the final steps common to all Arbs. Refer to Configuring Steps Common to all Arbs at
the end of this section.
Configuring Sinusoid Arbs
Step 1. Select the Arb Type and Shape. Power and Resistance types apply to .
Front Panel Menu Reference SCPI Command
Select Transient\Arb\Function.
Select a Voltage, Current, Power, or Resistance Arb Type.
Then selectthe Sine Shape. Then press Select.
To selecta current, voltage, power, or resistance sinusoid Arb,
use:
ARB:FUNC:TYPE VOLT|CURR|POW|RES,(@1)
ARB:FUNC:SHAP SIN,(@1)