EasyManua.ls Logo

Keysight N6700C Series - Page 193

Keysight N6700C Series
430 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Parameter Typical Return
Step 0 - 99, *RST 0 (none)
Programs a time constant for the current Arb: ARB:CURR:EXP:TCON 5,(@1)
Programs a time constant for the voltage Arb: ARB:VOLT:EXP:TCON 5,(@1)
Programs a time constant for the power Arb: ARB:POW:EXP:TCON 5,(@1)
Programs a time constant for the resistance Arb: ARB:RESISTANCE:EXP:TCON 5,(@1)
Programs a time constant for step 0: ARB:SEQ:STEP:CURR:EXP:TCON 5, 0,(@1)
[SOURce:]ARB:CURRent:EXPonential:TIMe <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:EXPonential:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:EXPonential:TIMe <value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:EXPonential:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:POWer:EXPonential:TIMe <value>, (@<chanlist>)
[SOURce:]ARB:POWer:EXPonential:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:RESISTANCE:EXPonential:TIMe <value>, (@<chanlist>)
[SOURce:]ARB:RESISTANCE:EXPonential:TIMe? [MIN|MAX,] (@<chanlist>)
Specifies the total time for the current, voltage, power, or resistance to go from the starting level to
the ending level in seconds.
[SOURce:]ARB:SEQuence:STEP:CURRent:EXPonential:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:EXPonential:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:EXPonential:TIMe<value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:EXPonential:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:EXPonential:TIMe<value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:EXPonential:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:EXPonential:TIMe<value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:EXPonential:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the total time for the current, voltage, power, or resistance to go from the starting level to
the ending level in seconds for Arb sequence steps.
Parameter Typical Return
0 262.144 | MIN | MAX, *RST 0 <total time>
Step 0 - 99, *RST 0 (none)
Programs the total time for the current Arb: ARB:CURR:EXP:TIM 10,(@1)
Programs the total time for the voltage Arb: ARB:VOLT:EXP:TIM 10,(@1)
Programs the total time for the power Arb: ARB:POW:EXP:TIM 10,(@1)
Programs the total time for the resistance Arb: ARB:RESISTANCE:EXP:TIM 10,(@1)
Programs the total time for step 0: ARB:SEQ:STEP:CURR:EXP:TIM 10, 0,(@1)
Keysight N6700C Operating and Service Guide 193
5 SCPI Programming Reference