Test Equipment and Setups
Equipment Required
Measurement Setups
Equipment Required
The test equipment recommended for the performance verification and adjustment procedures is
listed below. If the exact instrument is not available, substitute calibration standards of equivalent
accuracy.
Instrument Requirements Recommended Model Use
1
Digital Multimeter Resolution: 10 nV @ 1V
Readout: 8 1/2 digits
Accuracy: 20 ppm
Keysight 3458A or equivalent
(2 required for N679xA verification)
V, C
Current Shunt
2
15 A (0.1Ω )
30 A (0.1Ω )
50 A (0.05Ω )
Guildline 9230A-15R
Guildline 9230A-30
Guildline 9230A-50
V, C
Load Resistors 25 Ω , 250 W, 1% (N6751A/52A/61A/62A)
4 Ω , 250 W, 1% (N6751A/52A/61A/62A)
20 kΩ, 1%, 1/8W (for option 1UA/2UA)
10 Ω , 1%, 1/8W(N678xA SMU)
10 kΩ, 1%, 1/8W (N678xA SMU)
1 MΩ , 1%, 1/8W (N678xA SMU)
Vishay NH-250-25Ω -1%
Vishay NH-250-4Ω -1%
C
C
V, C
V, C
V, C
V, C
Electronic load
2
150 V, 50 A minimum, with transient capability and a
slewrate of 833 kA/s or better
Keysight N3300A mainframe, with
N330xA modules
V, C
Keysight N6784A Load for N678xA SMU and N6783A
(setto CC load mode)
Keysight N6784A V
GPIB controller Full GPIB capabilities Keysight 82350B or equivalent V, C
Oscilloscope Sensitivity: 1 mV
Bandwidth Limit: 20 MHz
Probe: 1:1 with RF tip
Keysight DSO6054A or equivalent V
RMS Voltmeter True RMS
Bandwidth: 20 MHz
Sensitivity: 100 µV
Rhode and Schwartz Model URE3 or
equivalent
V
Differential Amplifier Bandwidth: 20 MHz LeCroy DA1855A, DA1850A, or equi-
valent
V
Terminations 1 – 50 Ω BNC termination
2 – 50 Ω 1/8 W termination resistors
V
Power supply 80 V, 50 A, 300 Wminimum (for N679xA load modules) Keysight N6763A, N6764A, and
N6776A or equivalent
V, C