Auxiliary voltage measurements cannot be made along with output voltage measurements. When the
auxiliary voltage measurement input is selected, the voltage measurement input will be switched to
the Aux Voltage inputs instead of the normal + and − sense terminals.
To enable auxiliary voltage measurements:
Front Panel Menu Reference SCPI Command
Select Measure\Input.
Select Auxiliary. Then press Select.
Select Main to reconnect the voltage measurement
input to the output terminals.
To enable auxiliary measurements:
SENS:FUNC:VOLT:INP AUX,(@1)
Take the auxiliary voltage measurement:
MEAS:VOLT? (@1)
When the auxiliary voltage measurement terminals are left unconnected, the front
panel meter will indicate a voltage reading of approximately 1.6V. This is a normal
indication which does not affect the voltage measurement once the auxiliary
measurement terminals are connected.
Keysight N6700C Operating and Service Guide 133
4 Operating the Power System