EasyManua.ls Logo

Keysight N6700C Series - Page 228

Keysight N6700C Series
430 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Parameter Typical Return
DWELl | TRIGgered, *RST DWELl DWEL or TRIG
Step 0 - 99, *RST 0 (none)
Specify trigger pacing for sequence step 1: ARB:SEQ:STEP:PAC TRIG, 1, (@1)
l If the <step#> specified is an existing sequence step, that step will be changed to the specified
waveshape with all parameters reset to default values.
l New sequence steps must be specified sequentially. To create a new sequence step, <step#>
should be specified to be the current length of the sequence (see ARB:SEQuence:LENGth?). When
a step is added, all parameters of the step waveform are reset to their default values.
[SOURce:]ARB:SEQuence:TERMinate:LAST <Bool>, (@<chanlist>)
[SOURce:]ARB:SEQuence:TERMinate:LAST? (@<chanlist>)
Determines the output value when the Arb sequence terminates. The state is either ON (1) or OFF (0).
When ON The output remains at the last Arb sequencevalue. The last current, voltage, power, or resistance Arb value becomes the
IMMediate value when the sequence completes.
When OFF The output returns to the settings in effectbefore the sequence started. This also applies when the sequence is aborted,
Parameter Typical Return
0|OFF|1|ON, *RST Off 0or1
Sets the sequence to terminate with the voltage, resistance, or current setatthe last sequence value:
ARB:SEQ:TERM:LAST ON,(@1)
l The query returns 0 if the output returns to the settings that were in effect before the sequence
started, and 1 if the output remains at the last sequence value.
5 SCPI Programming Reference
228 Keysight N6700C Operating and Service Guide