CALibrate:SAVE
Saves the calibration constants in non-volatile memory. Do this at the end of the calibration to avoid
losing changes.
Parameter Typical Return
(none) (none)
Store cal constants in non-volatile memory: CAL:SAVE
l If calibration mode is exited by programming CALibration:STATe OFF without first saving the new
constants, the previous calibration constants are restored.
CALibrate:STATe 0|OFF|1|ON [,<password>]
CALibrate:STATe?
Enables/disables calibration mode. Calibration mode must be enabled for the instrument to accept
any calibration commands.
The optional parameter is the password. This is required if the existing password is not 0 (zero). If the
password is not entered or is incorrect, an error is generated and the calibration mode remains
disabled.
Parameter Typical Return
0|OFF|1|ON , default OFF 0or1
<password>a numeric value up to 15 digits (none)
Disable calibration mode and unsecure calibration: CAL:STAT OFF, 0
Enable calibration using a numeric password: CAL:STATON, 1234
l If calibration mode is exited by programming CALibration:STATe OFF without first saving the new
constants, the previous calibration constants are restored.
CALibrate:VOLTage[:LEVel] <value>, (@<channel>)
Calibrates the output voltage. The value entered selects the range that is being calibrated.
Parameter Typical Return
The maximum voltage of the range. (none)
Calibrates the voltage of the 50 V range: CAL:VOLT51, (@1)
CALibrate:VOLTage:CMRR (@<channel>)
Calibrates the voltage common mode rejection ratio.
5 SCPI Programming Reference
232 Keysight N6700C Operating and Service Guide