MEASure:ARRay:CURRent[:DC]? (@<chanlist>)
MEASure:ARRay:VOLTage[:DC]? (@<chanlist>)
MEASure:ARRay:POWer[:DC]? (@<chanlist>)
MEASure:ARRAy commands require to be installed on models N673xA.
N674xA, N675xA, and N677xA.
Initiates and triggers a measurement; returns a list of the digitized output measurement samples.
Values returned are either in amperes, volts, or watts.
The sampling rate is set by SENSe:SWEep:TINTerval. The position of the trigger relative to the
beginning of the data buffer is determined by SENSe:SWEep:OFFSet:POINts. The number of points
returned is set by SENSe:SWEep:POINts.
The return format depends on the settings of the FORMat:BORDer and FORMat[:DATA] commands.
When the data format is set to ASCII, returned values are comma separated. When the data format is
set to REAL, data is returned as single precision floating point values in definite length arbitrary block
response format.
Parameter Typical Return
(none) <value> [,<value>]
or <Block>
Returns the measured current array MEAS:ARR:CURR? (@1)
Returns the measured power array MEAS:ARR:POW? (@1)
Returns the measured voltage array MEAS:ARR:VOLT? (@1)
Keysight N6700C Operating and Service Guide 263
5 SCPI Programming Reference