EasyManua.ls Logo

Keysight N6700C Series - Page 299

Keysight N6700C Series
430 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Trigger Source Description Trigger System
BUS Selects a remoteinterfacetrigger command. All trigger systems
CURRent<1-4> Selects the output current level of a channel. Acquire triggers only
EXTernal Selects all digital port pins thathave been configured as trigger inputs. All trigger systems
IMMediate Generates a trigger as soon as the trigger system is INITiated. All trigger systems
PIN<1-7> Selects a digital port pin configured as a trigger input. All trigger systems except ARB
TRANsient<1-4> Selects the transient system as the trigger source. Acquire and Transient triggers
only
VOLTage<1-4> Selects the output voltage level of a channel. Acquire triggers only
Parameter Typical Return
BUS|CURRent<1-4>|EXTernal|PIN<1-7>| TRANsient<1-
4>|VOLTage<1-4>, *RST BUS
BUS, CURR<n>, EXT, PIN<n>, TRAN<n>, or VOLT<n>
Selects a current level on channel 1 as the acquire trigger:TRIG:ACQ:SOUR CURR1, (@1)
Selects a digital port pin 1 as the Elog trigger source: TRIG:ELOG:SOUR PIN1, (@1)
Selects the transient system of channel 4 as the trigger source: TRIG:TRAN:SOUR TRAN4, (@1)
Selects the voltage level of channel 3 as the acquisition trigger: TRIG:ACQ:SOUR VOLT3, (@1)
TRIGger:MEASure:TALign:CORRection[:STATe] 0|OFF|1|ON, (@chanlist)
TRIGger:MEASure:TALign:CORRection[:STATe]? (@chanlist)
Improves the time alignment of measurement triggers with measurement data when enabled. The
power-on state is OFF for backwards compatibility.
Parameter Typical Return
0|OFF|1|ON, *RST OFF 0or1
To improve the measurement trigger alignment: TRIG:MEAS:TAL:CORR ON, (@1)
Keysight N6700C Operating and Service Guide 299
5 SCPI Programming Reference

Table of Contents