Trigger Source Description Trigger System
BUS Selects a remoteinterfacetrigger command. All trigger systems
CURRent<1-4> Selects the output current level of a channel. Acquire triggers only
EXTernal Selects all digital port pins thathave been configured as trigger inputs. All trigger systems
IMMediate Generates a trigger as soon as the trigger system is INITiated. All trigger systems
PIN<1-7> Selects a digital port pin configured as a trigger input. All trigger systems except ARB
TRANsient<1-4> Selects the transient system as the trigger source. Acquire and Transient triggers
only
VOLTage<1-4> Selects the output voltage level of a channel. Acquire triggers only
Parameter Typical Return
BUS|CURRent<1-4>|EXTernal|PIN<1-7>| TRANsient<1-
4>|VOLTage<1-4>, *RST BUS
BUS, CURR<n>, EXT, PIN<n>, TRAN<n>, or VOLT<n>
Selects a current level on channel 1 as the acquire trigger:TRIG:ACQ:SOUR CURR1, (@1)
Selects a digital port pin 1 as the Elog trigger source: TRIG:ELOG:SOUR PIN1, (@1)
Selects the transient system of channel 4 as the trigger source: TRIG:TRAN:SOUR TRAN4, (@1)
Selects the voltage level of channel 3 as the acquisition trigger: TRIG:ACQ:SOUR VOLT3, (@1)
TRIGger:MEASure:TALign:CORRection[:STATe] 0|OFF|1|ON, (@chanlist)
TRIGger:MEASure:TALign:CORRection[:STATe]? (@chanlist)
Improves the time alignment of measurement triggers with measurement data when enabled. The
power-on state is OFF for backwards compatibility.
Parameter Typical Return
0|OFF|1|ON, *RST OFF 0or1
To improve the measurement trigger alignment: TRIG:MEAS:TAL:CORR ON, (@1)
Keysight N6700C Operating and Service Guide 299
5 SCPI Programming Reference