EasyManua.ls Logo

Keysight N6751A - Page 13

Keysight N6751A
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
N67xx DUT Power Supply Installation 1-9
Connecting to the ASRU Card
Connecting to Channels 1 through 4
The N6751 is a 4-output DUT power supply. One ASRU channel is required for each
output.
1 Locate the highest numbered module with channels 1 through 4 available on
the ASRU Card.
If channels 1 through 4 are not available, see Connecting to Channels 5 and 6
on page 1-10.
2 Connect the DUT Power Supply cable to the ASRU Card and to the N6751
power supply, as illustrated in Figure 1-5.
Figure 1-5 N6751 output connections – Channels 1 through 4
Channel 4 Channel 3 Channel 2 Channel 1
P/S Label on
cable
Cable color P/S Label on
cable
Cable color P/S Label on
cable
Cable color P/S Label on
cable
Cable color
+S F1 GRN +S G1 BLU +S H1 RED +S J1 BLK
+ F2 RED
RED/WHT
+ G2 BLU/WHT + H2 RED/WHT + J2 BLK/WHT
G3 GRN/WHT H3 ORN/WHT J3 BRN/WHT
F3 BLK
BLK/WHT
–S G4 WHT –S H4 WHT –S J4 WHT
–S F4 WHT
F1 F2 F3 F4
G1 G2 G3 G4
H1 H2 H3 H4
J1 J2 J3 J4