2-4 N67xx DUT Power Supply Installation
Replacing Connectors on Old Output Cables
The output cables from some older i3070 systems can be used with the N67xx DUT
power supplies by cutting off the lugs and rewiring them based on the diagrams in
Connecting the Output Cables.
Note that the labels on new output cables are different from the old cables, and the
wiring information in this chapter refer to the new labels. Table 2-1 refers you to the
correct tables for the mapping of old to new labels, depending on the output cable
and channel you are using.
Table 2-1
Refer to: Cable Channel
Table 2-2 E4000-61602 for 4-module system Channels 1–4
E1170-61607 2-module system Channels 1–4
E1170-61606 for 2-module system Channels 5–6
Table 2-3 E4000-61606 for 2-module system Channels 5–6
Table 2-4 E4000-61602 for 4-module system Channels 5–6
Table 2-2 Mapping old and new labels on output cables
Label on Cable: Label on Cable:
P/S Old New Cable Color P/S Old New Cable Color
+S 5 J1 BLK +S 5 H1 RED
+ 8 J2 BLK/WHT + 8 H2 RED/WHT
7 J3 BRN/WHT 7 H3 ORN/WHT
–S 6 J4 WHT –S 6 H4 WHT
+S 1 G1 BLU +S 1 F1 GRN
+ 4 G2 BLU/WHT + 4 F2 RED, RED/WHT
3 G3 GRN/WHT 3 F3 BLK, BLK/WHT
–S 2 G4 WHT –S 2 F4 WHT
+S 1 K1 GRN +S 5 L1 BLU
+ 4 K2 BLK/WHT + 8 L2 RED/WHT
3 K3 BRN/WHT 7 L3 ORN/WHT
–S 2 K4 WHT –S 6 L4 WHT
Channel 1
Channel 2
Channel 3
Channel 4
Channel 5
Channel 6