1-18 N67xx DUT Power Supply Installation
Connecting Output Cables for N6746
Power from the N6746 power supply can be routed to any single module, but must
use either ASRU channel 5 or 6. It is recommended that you use ASRU channel 6
as a first choice when configuring this supply. Up to four power outputs can be
routed to any single module or they can be split across four modules. (Each output
powers one module.)
Connect the output cables as shown in Figure 1-12.
Figure 1-12 N6746 output connections
Channel 5 or 6
P/S Label on cable Cable color
+S L1 BLU
+ L2 RED/WHT
L3 ORN/WHT
–S L4 WHT
+S K1 GRN
+ K2 BLK/WHT
K3 BRN/WHT
–S K4 WHT
L1 L2 L3 L4
K1 K2 K3 K4