Connecting Output Cables for N6773
Connecting to the ASRU Card
Connecting to the Utility Card
Connecting to the ASRU Card
The N6773 DUT Power Supply is a one-output, high-current supply. It can only be
connected to channels 5 and 6 on the ASRU Card. Two ASRU channels are required
to handle the high current. Connect the supply to the highest numbered module
with channels 5 and 6 available. Use cable N1807-61621 for a 4-module system or
N1807-61616 for a 2-module system.
Connect the cables as shown in Figure 1-10.
Figure 1-10 N6773 output connections
Channels 5 & 6
P/S Label on cable Cable color
+S K1 GRN
+ K2 BLK/WHT
K3 BRN/WHT
–S K4 WHT
+S L1 BLU
+ L2 RED/WHT
L3 ORN/WHT
–S L4 WHT