Loading...
5-12 Electronic Calibration Modules Reference Guide
Specifications and Characteristics
Characteristic Performance
5-
N4431D (7-16)
The characteristic performance in the following table applies to N4431D (if any of
the ports consist of 7-16 connector
). The data describes performance when
measuring “thru p
ath” A-B, C-D, A–D, A-D, B–C, and B-D.
N4431D (Type-N 50 Ω)
The characteristic performance in the following table applies to N4431D
Option 020 (Type-N 50 Ω female connectors on all ports). The data describes
performance when measuring “thru path” A-B, C-D, A–D, A-D, B–C, and B-D.
Table 5-21 Characteristic Performance for N4431D (m/f 7–16 connectors)
a
Parameter Frequency Range
DC to 10 MHz 10 MHz to 5 GHz 5 GHz to 7.5 GHz
Transmission
Tracking (±dB)
b
0.18 0.05 0.08
Load Match (dB)
b
35 42 40
a. When applied power exceeds –7 dBm, calibration results will be degraded from the performance indicated in this
table.
b. Values based on using the network analyzer N5232B 400.
Table 5-22 Characteristic Performance for N4431D (Type-N 50 Ω) Option 020
a
Parameter Frequency Range
DC to 10 MHz 10 MHz to 5 GHz 5 GHz to 9 GHz 9 GHz to 13.5 GHz
Transmission
Tracking (±dB)
b
0.18 0.05 0.08 0.12
Load Match (dB)
b
35 43 41 38
a. When applied power exceeds –7 dBm, calibration results will be degraded from the performance indicated in this
table.
b. Values based on using the network analyzer N5232B 400.