3-6 Electronic Calibration Modules Reference Guide
Operating ECal Module
Setting Up a Calibration
3-
Calibration Types Available
The following table contains the calibration types and advanced settings
available for all compatible network analyzers:
Table 3-3 Calibration Types and Advanced Settings Available
Calibration Types Model
PNA PXIe USB-VNA ENA
FieldFox
a
1-Port Reflection Yes Yes Yes Yes Yes
Full 2-Port Yes Yes Yes Yes Yes
Full N-Port where N>2 Yes Yes Yes Yes No
Enhanced Response Yes Yes Yes Yes Yes
QSOLT Yes Yes Yes No Yes
Advanced Settings
Adapter Removal Yes Yes Yes No No
Flush (zero length) THRU Yes Yes Yes No Yes
Unknown THRU Yes Yes Yes Yes Yes
User-Characterization Yes Yes Yes Yes
Yes
b
a. ECal calibrations are not supported on the N9912A.
b. FieldFox user-characterizations must be performed on a PNA or ENA. Learn more about User
Characterization at the PNA Help website: http://na.support.keysight.com/pna/help/latest/S3_-
Cals/ECal_User_Characterization.htm