Option Esc External Source Control
Using Option ESC with the Keysight MXG Signal Sources.
Down loading the conversion loss .csv files to the analyzer corrections array
Action Notes
1 Install the CD ROM provided 
with the mixer, into a PC.
You can view the contents of the CD.
2 Locate the 70xxxxxx_X.csv file. 
3 Copy the .csv file to the USB 
memory stick provided with the 
mixer.
4 Insert the USB memory stick into 
one of the USB ports on the signal 
analyzer.
It is recommended that you connect a 
mouse and keyboard to the signal analyzer.
5 Press Input/Output, More, 
Corrections, Select Correction.
6 Choose a correction array from 
the list of Correction 1 through 
Correction 6.
Correction 1 has a provision to store 
antenna corrections, so if antenna 
corrections are required, reserve this array 
for that use. 
7 To see if anything is already 
stored in a particular correction, 
press Correction, Edit.
Selecting Edit turns the correction ON. Be 
sure to turn the correction OFF after 
determining the contents of the correction 
array.
8 To delete a correction table, press 
Return, assure the Select 
Correction key corresponds to the 
correction you want to delete, and 
press Delete Correction.
9 Once a correction array number is 
selected, press Recall, Data 
(Import), Amplitude 
Correction, and choose one of the 
correction array numbers.
10Press Open, use the pull down 
arrow in the "look in" box to 
navigate to the USB memory 
stick, and locate the 
70xxxxxx_X.csv file.
11 Select the 70xxxxxx_X.csv file 
and click Open.
The conversion loss data will now load into 