214 
Concepts
Time Gating Concepts
Noise measurements using Time Gating
Time gating can be used to measure many types of signals. Noise and noise-like 
signals are often a special case in spectrum analysis. With the history of gated 
measurements, these signals are especially noteworthy.
The average detector is the best detector to use for measuring noise-like signals 
because it uses all the available noise power all the time in its measurement. The 
sample detector is also a good choice because it, too, is free from the peak biases of 
the peak detector, normal and negative peak detectors. 
When using the average or sample detector, noise density measurements using the 
noise marker or band/interval density marker can be made without any consideration 
of the use of gating--gated measurements work just as well as non-gated 
measurements. Thus, the average detector is recommended for noise density 
measurements.
Older analyzers only had the gated video version of gating available, and these only 
worked with the peak detector, so the rest of this section will discuss the trade-offs 
associated with trying to replicate these measurements with an X-Series analyzer.
Unlike older analyzers, X-Series analyzers can make competent measurements of 
noise density using the noise marker with all detectors, not just those that are ideal 
for noise measurements. Thus, X-Series analyzers can make noise density 
measurements with peak detection, compensating for the extent to which peak 
detection increases the average response of the analyzer to noise. When comparing a 
gated video measurement using the noise marker between an X-Series and an older 
analyzer where both use the peak detector, the X-Series answer will be 
approximately correct, while the older analyzer will need a correction factor. That 
correction factor is discussed in Keysight Technologies Application Note 1303, 
Spectrum Analyzer Measurements and Noise, in the section on Peak-detected Noise 
and TDMA ACP Measurements.
When making measurements of Band/Interval Power or Band/Interval Density, the 
analyzer does not make compensations for peak detection. For best measurements 
with these marker functions, average or sample detection should be used.