Measuring Noise
Improving Phase Noise Measurements by Subtracting Signal Analyzer Noise
Figure 7-11  Measurement of DUT and Analyzer Noise
7 Measure only the analyzer 
noise using trace 2 (blue 
trace) with trace averaging.
a. Turn off or remove the DUT signal to 
the RF input of the analyzer.
b. Press Trace/Detector, Select Trace, 
Trace 2, Clear Write, Trace 
Average.
c. Press View/Blank, View.
 
Allow time for sufficient 
averaging.
See Figure 7-12.
Step Action Notes