1 Getting Started Tutorial
32 Keysight U3402A User’s and Service Guide
Table 1-3 Input terminal for different measurement functions
Measurement function Input terminal Overload protection
DC voltage (VDC)
V Ω Hz (Hi)
Lo
1000 VDC
AC voltage (VAC), frequency (Hz)
750 VAC RMS, 1100 V PEAK,
2x10
7
V-Hz normal mode, or
1x10
6
V-Hz common mode
Miliampere (mA), frequency (Hz) mA 1200 mADC or AC RMS
12A, frequency (Hz) 12A
12 ADC or AC RMS for continuous
30 seconds, or 10 ADC or AC RMS
2-wire resistance (Ω (2W))
V Ω Hz 500 VDC or AC RMS
Diode test, continuity test
4-wire resistance (Ω (4W)) Hi 250 VDC or AC RMS
All functions Any terminal to earth 1000 VDC or AC PEAK