Performance Test 4
Keysight U3402A User’s and Service Guide 81
DC Current Verification Test
1 Connect the calibrator to the front panel Hi and Lo input connectors.
2 Select each function and range in the order shown in Table 4-3. Provide the
input shown in Table 4-3.
3 Make a measurement and observe the result. Compare measurement results
to the appropriate test limits shown in Table 4-3. (Be certain to allow for
appropriate source settling when using the Fluke 5520A.)
Table 4-3 DC current verification test
Function Reading rate Input Range
Error from nominal one
year
DC current Slow 0.0000 mA 12 mA ±1.5 µA
0.000 mA 120 mA ±5 µA
0.00 mA 1200 mA ±50 µA
0.0000 A 12 A ±0.5 mA
Medium 0.000 mA 40 mA ±6 µA
0.00 mA 120 mA ±30 µA
0.0 mA 1200 mA ±0.3 mA
0.000 A 12 A ±3 mA
Slow 10.0000 mA 12 mA ±6.5 µA
100.000 mA 120 mA ±55 µA
1000.00 mA 1200 mA ±2.05 mA
10.0000 A 12 A ±20.5 mA
Medium 36.000 mA 40 mA ±42 µA
100.00 mA 120 mA ±0.13 mA
1000.0 mA 1200 mA ±2.3 mA
10.000 A 12 A ±23 mA
CAUTION
Connect calibrator to multimeter’s A and Lo terminals before applying 10 A.