Operations and Features 2
Keysight U3402A User’s and Service Guide 55
Procedure
1 Press to enable Rel operation.
2 Press
to toggle in (and out) the Rel #.
3 Use , , , and to modify the Rel #.
4 Press to set the Rel # and read the display.
5 Press to disable the Rel operation.
MinMax
The MinMax (Minimum/Maximum) operation stores the minimum and maximum
values of reading during a series of measurements.
When enabled, the MinMax operation turns on the MINMAX annunciator and
begins accumulating various statistics of the readings being displayed.
Figure 2-3 Typical Max operation display
NOTE
In resistance measurement mode, the multimeter will read a non-zero value
even when the two test leads are in direct contact, because of the resistance
of these leads. Use the Rel operation to zero-adjust the display.
In DC voltage measurement mode, the thermal effect will influence the
accuracy. Short the test leads and press once the displayed value is
stable to zero-adjust the display.