Application Tutorial 3
Keysight U3402A User’s and Service Guide 67
Dual Display Operation Examples
This section describes some practical operations when using dual display feature.
Measure DC Voltage and AC Ripple on a Rectification Circuit
A single measurement for both DC voltage and AC ripple can be displayed
through both display while testing a rectifier circuit.
1 Connect the red and black test leads to the input terminal and probe the test
points as shown in Figure 3-1.
Figure 3-1 Terminal connection when measuring DC voltage and AC ripple
on a rectification circuit
2 Press to select DC voltage measurement for primary display.
3 Press to enable the secondary display. The 2ND is displayed.
4 Press to select AC voltage measurement for secondary display.
AC/DC
voltage
source
+