Calibration Procedures 1
Keysight U3606B Service Guide 35
Figure 1-12 Test setup for constant voltage noise effect (peak-to-peak and
RMS noise) verification
Figure 1-13 Test setup for load transient response time verification
+
–V
+V
V
out
V
1
V
2
R
1
R
1
R
2
R
2
4321
Fixed
resistor
BNC
BNC
Differential amplifier
BNC
Oscilloscope
50
Ω
termination
50
Ω
termination
RMS voltmeter
4321
LOAD
BNC
Oscilloscope
Electronic load

Table of Contents